HME Filter (FOR USE WITH VENITILATOR/RESPIRATOR)

Specifications:

  • Recommended Tidal Volume: 250ml ~ 1500ml
  • Temperature (ISO 9360)
    • Vt 250ml: 32°C
    • Vt 500ml: 32°C
    • Vt 750ml: 32°C
    • Vt 1000ml: 32°C
  • Absolute Humidity (500ml Vt, 20BPM): 30.6mg H2O/L
  • Dead Space: 45ml
  • Bacterial Filtration Efficiency: 99.99674%
  • Virus Filtration Efficiency: 99.99362%
  • Resistance to Flow (ISO 9360)
    • 15/min, 30/min, 60/min:<1.0cm H2O, <2.5cm H2O, 3.5cm H2O
  • Filtration Media: Electrostatic
  • Sterile: Gamma/E.O Gas

Features:

  • Combined Filter/HME reduces set-up time, minimizes dead space and weight on endotracheal tube and circuit while maximizing efficiency.
  • Electrostatic Hydrophobic membrane protects against cross-contamination.
  • Hygroscopic membrane traps patient’s moisture providing effective airway humidification.
  • Low resistance to airflow minimizes work of breathing.
  • Effective moisture control reduces the risk of colonization in breathing tubes and minimizes sampling line blockage.
  • End tidal CO2 sampling port offers convenient access to airway gases.
  • ISO standard 15mm and 22mm fitting connects with breathing system.
  • Treated with physiologically safe calcium chloride.
  • Patient/Machine end.

Certificate:

  • FDA
  • EN46002
  • ISO 9001

 

 

Comments are closed.